Detector Unclog -Detector Debubble -Detector Long Clean
– Detector Unclog – Detector Debubble – Detector Quick Clean Perform Fluidics Maintenance if it has been close to or over 176 hours of runtime since it was previously performed.Replace the probe and tubing followed by a probe and tubing clean.We recommend this be performed after 176 hours of runtime or every 2 months. Fluidics Maintenance is a procedure in which the in-line sheath filter, fluidlink and bottle filters are changed out.The probe and tubing can become degraded or clogged after a time, we recommend it be changed after 20 hours of use.Beads should be prepared by adding 2 drops of beads to 300 uL of Buffer Solution (containing 1% PBS and 0.1% BSA) in a 1.5 mL microcenterfuge tube and placed in a vortexer just prior to sampling.We recommend using a bead prep no longer than 5 days and should be stored at 2-8 degrees C.The Low Event Count metric is used to determine the number of events counted during a QC test.